PSA nitrogen generator NITROSource Plus
Produces nitrogen from compressed air. Nitrogen provides a clean, dry and inert atmosphere wherever oxygen is unwanted in environments and processes.
NSP-020 – NSP-120 · 11 models
- Nitrogen outlet flow
- 7.0 – 197.3m³/h
- At 7 bar(g) inlet, 20 °C ambient
- Purity · residual oxygen
- 5 ppm – 3%
- Flow decreases as purity increases
- Operating pressure
- 5 – 13bar(g)
- Compressed air inlet
- Series
- 11models
- Runs in parallel as multi-bank
Air is 78% nitrogen; stop paying for gas.
NITROSource Plus is Parker’s new series of nitrogen generators based on pressure swing adsorption (PSA). Holding residual oxygen between 3% and 5 ppm, it produces nitrogen at the purity each process needs. The manufacturer has had the series independently tested by an accredited body for compliance with global standards for food, beverage and pharmaceutical-grade nitrogen; its carbon molecular sieve (CMS) beds are designed for a service life of more than 10 years.
As a cost-effective solution, a Parker domnick hunter nitrogen generator pays back quickly. On-site generation also removes the burdens of external supply such as refills, ordering, tank rental, delivery charges and future price increases by the supplier.
The nitrogen generator can be fed from a standard air supply; on-demand delivery and economy mode let you control nitrogen production, flow and purity. The nitrogen produced is not wasted as with conventional sources, and the evaporation losses of liquid nitrogen disappear, bringing extra savings. Running 24/7 without interruption, it prevents downtime and production losses caused by running out of nitrogen.

Because Parker NITROSource nitrogen generators produce only the nitrogen you need, they remove all the hazards and costs of high-pressure cylinders.
The nitrogen generation system is easily installed in the compressor room or production area with standard piping and no special equipment. Maintenance is very short, so there is no downtime from gas interruptions. Nitrogen is produced at low pressure, avoiding the hazards of storing gas in high-pressure cylinders, and the labour of changing and moving cylinders disappears.
Multi-modular system
Thanks to its modular concept, the NITROSource nitrogen generator is more flexible than twin-tower PSA generators. Growing nitrogen demand is easily met by adding modules. Added modules provide redundancy, so there is no gas interruption during maintenance. Its compact design even fits through standard doorways.
Advantages and features.
- 01
Up to 90% cost savings*
The investment pays back in approximately 12–24 months.
- 02
Energy savings
Low air consumption brings substantial energy savings. Energy Saving Technology (EST) extends the column changeover time at partial load, reducing air consumption.
- 03
Practical and reliable
Simple to install, easy to use and low on maintenance costs. It removes the safety risks of other gas sources.
- 04
Space-saving design
Its compact design takes up less space in the plant.
- 05
Easy capacity expansion
The modular design easily meets growing nitrogen demand by adding extra columns.
- 06
Low carbon emissions
It avoids the environmental and air pollution caused by delivery vehicles in liquid and cylinder nitrogen supply.
* Calculated in comparison with cylinder or liquid nitrogen supply.
Features
- Can run on the plant’s existing compressed air supply.
- Reduces oxygen content to the required level between 5 ppm and 3%.
- Automatic economy mode (Energy Saving Technology).
- Integrated oxygen analyser for continuous purity monitoring: zirconia as standard; electrochemical option for pharmaceutical applications.
- Off-spec gas is vented automatically before it reaches the application.
- A mass flow controller and outlet pressure regulation keep nitrogen flow and pressure constant.
- Digital and analogue outputs for remote monitoring and control: web server, Modbus TCP/IP, 4–20 mA.
- Alarm functions: warning and fault relays, alarm and service logs.
- Easy-to-use control panel: 7-inch colour touchscreen, data logging and graphs.
- Compact design.
The real cost of cylinder supply
- 4,500
- compressed gas cylinders
- >100hours
- order management
- >100hours
- receiving, checking and moving cylinders
- 2,500miles
- delivery distance
- 3.5tonnes
- CO₂ from deliveries
- 10–15%
- of gas returned unused
Example: about 10 m³/h of nitrogen supplied in cylinders or bundles for 4,000 hours a year.
Two column banks, uninterrupted nitrogen.
In PSA nitrogen generators the vessels are filled with CMS (carbon molecular sieve) granules, a special adsorbent. Select the steps to follow the cycle.
Model, purity and flow.
For the same model, nitrogen flow drops as purity increases. Let us choose the right model together based on the purity you need and your consumption profile.
| Model | Residual oxygen· nitrogen flow, m³/h | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| 5 ppm | 10 ppm | 50 ppm | 100 ppm | 250 ppm | 500 ppm | 0.1% | 0.5% | 1% | 2% | 3% | |
| NSP-020 | 7.0 | 8.0 | 10.1 | 11.4 | 13.4 | 15.1 | 17.0 | 23.2 | 27.3 | 32.8 | 37.5 |
| NSP-030 | 9.4 | 10.8 | 14.6 | 16.6 | 19.7 | 23.1 | 25.8 | 34.7 | 40.9 | 48.7 | 55.1 |
| NSP-040 | 12.5 | 14.4 | 19.4 | 22.1 | 26.2 | 30.8 | 34.4 | 46.3 | 54.6 | 64.9 | 73.4 |
| NSP-050 | 15.6 | 18.0 | 24.3 | 27.7 | 32.8 | 38.5 | 43.0 | 57.9 | 68.2 | 81.1 | 91.8 |
| NSP-060 | 18.8 | 21.5 | 28.6 | 32.5 | 38.9 | 44.3 | 50.0 | 68.1 | 80.2 | 95.4 | 110.1 |
| NSP-070 | 21.9 | 25.1 | 33.3 | 38.0 | 45.4 | 51.7 | 58.3 | 79.4 | 93.6 | 111.3 | 128.5 |
| NSP-080 | 25.0 | 28.7 | 38.1 | 43.4 | 51.9 | 59.1 | 66.7 | 90.8 | 106.9 | 127.2 | 146.8 |
| NSP-090 | 28.1 | 32.3 | 42.8 | 48.8 | 58.4 | 66.4 | 75.0 | 102.1 | 120.3 | 143.1 | 165.2 |
| NSP-100 | 34.3 | 39.1 | 52.0 | 56.8 | 66.1 | 74.1 | 83.8 | 114.8 | 134.7 | 164.5 | 178.7 |
| NSP-110 | 37.7 | 43.0 | 57.2 | 62.5 | 72.7 | 81.5 | 92.2 | 126.3 | 148.2 | 180.9 | 196.2 |
| NSP-120 | 41.2 | 46.9 | 62.4 | 68.2 | 79.3 | 88.9 | 100.6 | 137.8 | 161.6 | 197.3 | — |
Flow rates are for 7 bar(g) inlet pressure and 20 °C ambient temperature; m³ refers to 20 °C, 1013 mbar(a) and 0% relative humidity.
| Parameter | Value | Unit |
|---|---|---|
| Minimum inlet air quality | ISO 8573-1:2010 · 2.4.1 | class |
| Operating pressure | 5 – 13 | bar(g) |
| Design pressure | 15 | bar(g) |
| Operating temperature | 5 – 50 · recommended 20 | °C |
| Supply voltage | 100 – 240 (±10%) · 50/60 | V AC · Hz |
| Power consumption | 100 | W |
| Protection class | IP32 · NEMA 2 | — |
| Noise level | < 75 | dB(A) |
A buffer tank smaller than recommended may cause outlet pressure and purity to fluctuate. We size it during the site survey.
| Model | Height mm | Width mm | Depth mm | Weight kg | Air inlet BSPP | Nitrogen outlet BSPP | Min. buffer tank L |
|---|---|---|---|---|---|---|---|
| NSP-020 | 2,063 | 550 | 883 | 335 | 1" | 1" | 270 |
| NSP-030 | 2,063 | 550 | 1,052 | 429 | 1" | 1" | 500 |
| NSP-040 | 2,063 | 550 | 1,221 | 522 | 1" | 1" | 500 |
| NSP-050 | 2,063 | 550 | 1,390 | 615 | 1" | 1" | 1,000 |
| NSP-060 | 2,063 | 550 | 1,559 | 710 | 1" | 1" | 1,000 |
| NSP-070 | 2,063 | 550 | 1,728 | 805 | 1" | 1" | 1,000 |
| NSP-080 | 2,063 | 550 | 1,897 | 897 | 1" | 1" | 1,500 |
| NSP-090 | 2,063 | 550 | 2,028 | 988 | 1" | 1" | 1,500 |
| NSP-100 | 2,063 | 550 | 2,194 | 1,104 | 1½" | 1" | 1,500 |
| NSP-110 | 2,063 | 550 | 2,360 | 1,197 | 1½" | 1" | 1,500 |
| NSP-120 | 2,063 | 550 | 2,526 | 1,292 | 1½" | 1" | 2,000 |
The buffer tank is the smallest volume recommended to achieve maximum outlet pressure.
Where it is used.
Brochure and information notes.
- BrochureNITROSource Plus PSA nitrogen generators
- InfographicOn-site nitrogen generation
- InfographicDrawbacks of cylinder supply
- Information noteWhat does nitrogen purity mean?
- Information notePharmaceutical-grade nitrogen
- Information noteWhat is a buffer tank?
- Information noteMass flow controller
- Information noteReference units for flow and volume
- Information noteCMS filling method
The team that installs it also maintains it.
It is the DHE team that selects, installs and commissions NITROSource Plus. We also handle periodic maintenance, breakdown response and spare parts ourselves, without handing them over to a dealer.
- Site survey and sizingWe measure your consumption, purity and air line on site.
- Installation and commissioningThe whole line, from compressor to point of use.
- Periodic maintenanceFilter elements, analyser checks, service records.
- Genuine spare partsFrequently replaced parts in stock.
About NITROSource Plus.
What should be considered when choosing a nitrogen generator?
The user’s process information, consumption and consumption trend must be analysed carefully. Since each application needs different nitrogen purities and pressures, the aim is both to meet the requirement and to offer the most economical solution.
How does purity affect flow?
The same generator model can produce nitrogen at different purities, but the flow depends on purity: the less oxygen required, the lower the outlet flow. Defining the purity you need correctly lets you choose a smaller and therefore more economical system. With NITROSource Plus, residual oxygen can be set between 5 ppm and 3%.
What is the payback period?
According to Parker, the typical payback period is 12–24 months; savings of up to 90% are possible compared with cylinder or liquid supply. The actual figure for you depends on your consumption and current supply cost.
What compressed air quality does the generator need?
Inlet air must be at least ISO 8573-1:2010 Class 2.4.1; a dryer and filters are installed upstream of the generator for this. During the site survey we measure your existing line and add pre-treatment if needed.
What if our nitrogen demand grows?
You don’t need to replace the generator with a bigger one; a parallel module is added to the system. Added modules provide redundancy, so there is no gas interruption during maintenance.
Who carries out maintenance?
DHE’s own technical service team. It handles periodic maintenance, analyser checks, filter element replacement and breakdown response; we keep frequently replaced parts in stock.
From the same family.
-
Parker · PSA nitrogen generator NITROSource Compact
A compact, lightweight and modular design for low and medium on-site nitrogen demand.View product -
Parker · Membrane nitrogen module HiFluxx
Membrane modules that meet high nitrogen flow demands from a standard compressed air supply at pressures up to 13 bar(g).View product
Let’s choose the right model together.